            Overview

            Product name: EasyStep Human NGAL(Neutrophil Gelatinase Associated Lipocalin) ELISA Kit
            Reactivity: Human
            Alternative Names: LCN2; p25; Lipocalin 2; Oncogene 24p3; 25 kDa alpha-2-microglobulin-related subunit of MMP-9; Siderocalin LCN2
            Assay Type: Sandwich
            Sensitivity: 5.6 pg/mL
            Standard: 1000 pg/mL
            Detection Range: 15.63-1000 pg/mL
            Sample Type: Serum, plasma
            Assay Length: 1.5h
            Research Area: Infection immunity;Urology;
            Uniprot ID: P80188
            Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human NGAL, and the Human NGAL standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human NGAL. After TMB substrate solution is added, only those wells that contain Human NGAL and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human NGAL in the samples is then determined by comparing the OD of the samples to the standard curve.

            標(biāo)準(zhǔn)曲線

            Concentration (pg/mL) OD Corrected OD
            1000.00 2.153 2.045
            500.00 1.515 1.407
            250.00 1.106 0.998
            125.00 0.928 0.820
            62.50 0.563 0.455
            31.25 0.426 0.318
            15.63 0.156 0.048
            0.00 0.108 0.000

            精密度

            Intra-assay Precision (Precision within an assay):CV%<8%

            Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

            Inter-assay Precision (Precision between assays):CV%<10%

            Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

            回收率

            Matrices listed below were spiked with certain level of recombinant NGAL and the recovery rates were calculated by comparing the measured value to the expected amount of NGAL in samples.
            Matrix Recovery range Average
            serum(n=5) 94-105% 98%
            EDTA plasma(n=5) 94-107% 100%
            Heparin plasma(n=5) 82-95% 88%

            線性

            The linearity of the kit was assayed by testing samples spiked with appropriate concentration of NGAL and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
            Matrix 1:2 1:4 1:8 1:16
            serum(n=5) 90-98% 85-92% 83-96% 93-102%
            EDTA plasma(n=5) 85-94% 87-96% 82-93% 89-103%
            Heparin plasma(n=5) 85-92% 83-101% 87-98% 79-92%