            Product name: Rat BMP2(Bone Morphogenetic Protein 2) ELISA Kit
            Reactivity: Rat
            Alternative Names: BMP2A; BMP-2A; Hemochromatosis Modifier
            Assay Type: Sandwich
            Sensitivity: 5.8 pg/mL
            Standard: 1000 pg/mL
            Detection Range: 15.63-1000 pg/mL
            Sample Type: Serum, plasma, tissue homogenates and other biological fluids
            Assay Length: 3.5h
            Research Area: Developmental science;Bone metabolism;
            Uniprot ID: P49001
            Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat BMP2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat BMP2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat BMP2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat BMP2 in the samples is then determined by comparing the OD of the samples to the standard curve.

            標(biāo)準(zhǔn)曲線

            Concentration (pg/mL) OD Corrected OD
            1000.00 2.207 2.125
            500.00 1.613 1.531
            250.00 1.088 1.006
            125.00 0.883 0.801
            62.50 0.427 0.345
            31.25 0.297 0.215
            15.63 0.165 0.083
            0.00 0.082 0.000

            精密度

            Intra-assay Precision (Precision within an assay):CV%<8%

            Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

            Inter-assay Precision (Precision between assays):CV%<10%

            Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

            回收率

            Matrices listed below were spiked with certain level of recombinant BMP2 and the recovery rates were calculated by comparing the measured value to the expected amount of BMP2 in samples.
            Matrix Recovery range Average
            serum(n=5) 92-102% 97%
            EDTA plasma(n=5) 81-95% 88%
            Heparin plasma(n=5) 95-103% 99%

            線性

            The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BMP2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
            Matrix 1:2 1:4 1:8 1:16
            serum(n=5) 87-98% 85-96% 92-102% 81-93%
            EDTA plasma(n=5) 86-98% 91-103% 85-97% 87-96%
            Heparin plasma(n=5) 87-90% 87-98% 85-96% 85-95%