            Product name: Human ProBDNF(Pro Brain Derived Neurotrophic Factor) ELISA Kit
            Reactivity: Human
            Alternative Names: Neurotrophin; Abrineurin
            Assay Type: Sandwich
            Sensitivity: 6.3 pg/mL
            Standard: 1000 pg/mL
            Detection Range: 15.63-1000 pg/mL
            Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
            Assay Length: 3.5h
            Research Area: Cytokine;Infection immunity;Neuro science;
            Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human ProBDNF. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human ProBDNF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human ProBDNF, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human ProBDNF in the samples is then determined by comparing the OD of the samples to the standard curve.

            標準曲線

            Concentration (pg/mL) OD Corrected OD
            1000.00 1.842 1.753
            500.00 1.507 1.418
            250.00 1.279 1.190
            125.00 0.887 0.798
            62.50 0.594 0.505
            31.25 0.355 0.266
            15.63 0.208 0.119
            0.00 0.089 0.000

            精密度

            Intra-assay Precision (Precision within an assay):CV%<8%

            Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

            Inter-assay Precision (Precision between assays):CV%<10%

            Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

            回收率

            Matrices listed below were spiked with certain level of recombinant ProBDNF and the recovery rates were calculated by comparing the measured value to the expected amount of ProBDNF in samples.
            Matrix Recovery range Average
            serum(n=5) 83-99% 95%
            EDTA plasma(n=5) 94-105% 98%
            Heparin plasma(n=5) 82-94% 88%

            線性

            The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ProBDNF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
            Matrix 1:2 1:4 1:8 1:16
            serum(n=5) 85-94% 87-101% 86-97% 88-95%
            EDTA plasma(n=5) 88-97% 90-99% 85-97% 79-101%
            Heparin plasma(n=5) 78-94% 97-105% 88-101% 82-96%